• Heater Tube Deposit Rater

Analytical Instrumentation

Heater Tube Deposit Rater

Mar 30 2011

The Jet Fuel Thermal Oxidation Test (JFTOT®) ASTM D 3241 is universally used by the industry to measure high temperature stability of aviation turbine fuels. The ASTM D3241 is required to be run on every batch of jet fuel produced according to ASTM D 1655 or DEF STAN 91-91 specifications. In this test method, the fuel is pumped through heater tube at fixed flow rate and during a specified period of time. The fuel is pass or fail rated according to the amount of deposit formed on the heater tube at specified temperature.

Traditionally the amount of deposit on the tube is rated visually by the operator against reference color scale. The proper rating requires significant experience and expertise. But, operator capabilities vary, so evaluation of colour can be quite subjective.

Many articles have been published on the fact that color does not provide real information on the thickness and volume of deposits, parameters which are far more meaningful for characterizing jet fuels for users and suppliers.
The aviation fuel industry found that one of the most crucial stage of this test method is an objective analysis of deposit produced on the heater tube.

A new instrumental method of quantitative measurement of tube deposits has been developed by AD Systems (France), in which the thickness of the deposit is accurately determined by an automatic instrument, reducing test subjectivity.

The innovative DR 10 instrument uses an interferometry technique for precise measurement of deposit thickness in nanometers range. This stand alone compact instrument can be easy installed at any location ideal for workflow. The operation is based on a powerful light source, a spectrometer with a fiber optic probe, and specially designed application software.

The DR 10 test is simple and straightforward. The heater tube is prepared according to ASTM D 3241 test procedure and is then placed in the test chamber of the DR10. Using intuitive graphical interface with touch screen panel, the operator enters sample information and starts a tube scan. Specific light is emitted on the surface of the heater tube. The reflected light is collected and the interference created by the deposit is monitored by the spectrometer. The built-in software analyzes the interference fringes and calculates the deposit thickness. The mechanical system displaces the optical probe and rotates the heater tube taking precise thickness measurements at 1,200 points along and around its surface. A detailed test report is ready in less than 15 minutes.

The DR 10 test report provides a 3D profile of the deposit distribution on the tube surface. The software automatically detects and reports the Standard Spot value which is mean deposit thickness of the thickest 2.5 mm2 area as defined in the ASTM D 3241. In addition, the average and the maximum thickness deposit values, as well as the calculated total deposit volume are reported. For complete test traceability, the DR 10 report is tagged with the heater tube serial number picture taken by the built-in camera.

The DR 10 is a useful measurement tool for fuel research programs providing unique information on fuel deposit behavior at different oxidation temperatures and for the evaluation of antifouling additives efficiency.
 


Digital Edition

PIN 25.2 Apr/May

May 2024

Safety - Carbon monoxide toxic and flammable gas detection Analytical Instrumentation - Density: A fundamental parameter at critical stages within the petroleum sector - Advancements and...

View all digital editions

Events

The World Biogas Expo 2024

Jul 10 2024 Birmingham, UK

Thailand Oil & Gas Roadshow 2024

Jul 11 2024 Rayong, Thailand

HPLC 2024

Jul 20 2024 Denver, CO, USA

ICMGP 2024

Jul 21 2024 Cape Town, South Africa

Colombia Oil & Gas

Jul 24 2024 Bogata, Colombia

View all events