• Organic Chloride in Crude by ASTM D4929 Part C

Analytical Instrumentation

Organic Chloride in Crude by ASTM D4929 Part C

Scope

ASTM D4929 is the Standard Test Method for Determination of Organic Chloride Content in Crude Oil.  Procedure C covers the measurement of organic chloride in the washed naphtha fraction of crude oil by X-ray fluorescence spectrometry.

The analysis of organic chloride in crude is demonstrated per ASTM D4929 Part C, whereby the peak intensity of the chlorine Kα line of the washed naphtha fraction of a crude oil specimen is measured by energy dispersive X-ray fluorescence (EDXRF) spectrometry or by monochromatic energy dispersive X-ray fluorescence (MEDXRF).

Background

Chlorides in crude contribute to corrosion in the piping at refineries during cracking as well as mid-stream in pipelines. Organic chlorides do not naturally occur in crude, however inorganic chlorides in the form of salts as well as residual organic chlorides from various natural sources as well as adulteration can contaminate crude. Contracts at the pipelines may contain clauses limiting the amount of organic chloride allowed in the crude. And at the refinery after desalting and desulfurisation, crude needs to be analysed for any residual organic chlorides possibly still entrained in the feedstock to avoid potential damage during the refining process.

Instrumentation

  1. Type: EDXRF
  2. Model: Rigaku NEX DE
  3. X-ray tube: 12W Ag-anode
  4. Detector: SDD
  5. Sample tray: 15-position Austosampler
  6. Film: Polypropylene (4um)
  7. Analysis Time: 300 sec
  8. Environment: Helium
  9.  
  10. Type: Monochromatic EDXRF (MEDXRF)
  11. 90â—¦ Cartesian Geometry Polarisation
  12. Model: Rigaku NEX CG
  13. X-ray tube: 50 W Pd-anode
  14. Detector: SDD
  15. Sample tray: 15-position Austosampler
  16. Film: Polypropylene (4um)
  17. Analysis Time: 300 sec
  18. Environment: Helium

Sample Preparation

ASTM D4929 is designed for the measurement of residual organic chlorides in crude. The crude sample is first prepared by distillation and wash to remove H2S and inorganic chlorides. After distillation and wash processes the resulting naphtha fraction is analyzed for Cl content by XRF using Part C methodology. The naphtha fraction is typically stable containing less than 1000 μg/g S. XRF calibration is made using mineral oil calibration standards, as mineral oil models naphtha for X-ray response.

To prepare a sample for XRF analysis simply place 4.0g of the naphtha sample in a standard 32mm XRF sample cup and measure directly.

ASTM D4929 Calibration - Part C XRF

Empirical calibration was made using commercially available mineral oil standards. For optimum calibration the standards contain both S and Cl, with S content ranging from 100 to 1000 μg/g. Alpha corrections are automatically employed to compensate for sulfur's effect on chlorine X-rays.

Calibration

Both sulfur and chorine are measured during calibration.  The sulfur measurement is used to optimize the measurement of the chlorine, to compensate for the effect of sulfur's absorption of chlorine X-rays.

Precision & Repeatability

To demonstrate recovery, precision and repeatability of the measurement, a few calibration standards were selected with results shown in the accompanying tables.

Detection Limits

The empirical method is used to determine detection limits using 300 sec measurement time. Ten repeat analyses of a blank mineral sample are taken with the sample in static position and the standard deviation (σ) is determined. The Lower Limit of Detection (LLD) is then defined as three times the standard deviation (3σ).

(See LLD tables)

Conclusion

The Rigaku NEX DE EDXRF spectrometer offers excellent performance for the measurement of organic chloride in crude per ASTM D4929 Part C. Compact, and suitable for use in the field or at refineries, the analyser also complies with many other ASTM and ISO methods for critical measurements in the petroleum industry, making it an excellent choice as a multi-purpose analyser in the petroleum industry. 

The Rigaku NEX CG elemental analyser, using 90º Cartesian Geometry polarisation monochromatic EDXRF, gives superior performance for the measurement of organic chloride in crude as per ASTM D4929 Part C. The NEX CG analyser also complies with many other ASTM methods and EPA requirements for critical measurements in the petroleum industry, making it an excellent choice as an all-in-one analyser for measuring sulfur and ultra-low sulfur, metals in crude and residual oil, and lube oils.


Digital Edition

PIN 25.5 Oct/Nov 2024

November 2024

Analytical Instrumentation - Picturing Viscosity – How Can a Viscometer or a Rheometer Benefit You? - Sustainable Grease Formulations: Evaluating Key Performance Parameters and Testing Method...

View all digital editions

Events

Valve World Expo

Dec 03 2024 Dusseldorf, Germany

ASTM - DO2 Meeting

Dec 08 2024 Anaheim, CA, USA

Turkey & Black Sea Oil and Gas

Dec 11 2024 Istanbul, Turkey

IGC 2024

Dec 19 2024 Aurangabad, India

Clean Fuels Conference

Jan 20 2025 San Diego, CA, USA

View all events