Safe & Fast Method for the Analysis of Peroxides in Hydrocarbon Streams
May 19 2020
Peroxides present a potential safety hazard during production and storage. They also have a negative effect on fuels and fuel systems by reducing the octane rating in gasoline. To monitor the trace amounts of peroxides Da Vinci Laboratory Solutions has developed a chromatographic solution based upon a flow injection analysis as a safe and fast alternative to the traditional analysis methods, such as ASTM D2340 and ASTM E299.
This solution has been introduced in 2016 as the DVLS Fast Peroxide Analyser (FPA). The FPA system is based upon an Agilent 1260 infinity II HPLC stack configured with a dedicated reaction oven module to automate the safe, fast and precise analysis of active oxygen originated from hydrogen peroxides and other organic peroxides.
Flow Injection Analysis
To solve the air interference issue of the traditional analysis methods a flow injection technique is used for the Fast Peroxide analysis. The analysis of Peroxide, as active Oxygen, is based on the oxidation of Iodide by Peroxide:
ROOH + 2I- + 2H → ROH + I2 + H2O
Two reagents: Acetic Acid in n-Propanol and Sodium Iodide in n-Propanol are mixed, resulting in a stream of acidified Iodine solution in n-Propanol. The sample is injected into this stream. Subsequently, the liquid stream is lead through a heated inert reaction coil. Iodine is formed under temperature controlled conditions inside the dedicated reaction oven developed by Da Vinci. After passing the reaction coil the HPLC UV detector spectrophotometrically determines the formed Iodine, which is proportional to the level of peroxide.
Dedicated Configuration
The DVLS Fast Peroxide Analyser is based upon the Agilent 1260 Infinity II Liquid Chromatograph (LC) configured with:
- A Quaternary Pump
- An Automatic Liquid Sampler
- An UV Detector for a single wavelength detection
- An DVLS Reaction Oven module to hold and heat the inert reaction coil (Figure Two)
The reaction oven developed by Da Vinci enables an optimal heat distribution to the reaction coil. The oven design matches the HPLC modules thus allowing stacking of the oven between the autosampler and the detector.
Wide Application Range
The Fast Peroxide Analyser was initially developed to monitor peroxides in alpha-olefinic streams, it can also be used to determine peroxides in different matrices such as:
- 1,3-Butadiene
- Aqueous process streams
- Glycols
- Styrene
Performance at ppb level
Linearity, detection limits and repeatability were determined by analysing a series of Hydrogen Peroxide dissolved in 1-Propanol at a concentration range of 5 up to 324 μg/L active Oxygen (Figure Three). Each concentration level was analysed in tenfold, resulting in an LOD of 9.52 ppb w and LOQ of 28.8 ppb w.
Key Benefits:
- Safe and automated alternative to traditional wet chemistry analysis methods
- Wide application range: from alpha-olefinic streams to 1,3-Butadiene, Aqueous process streams, Glycols and Styrene
- Excellent performance down to ppb level
- Configuration based on the standard Agilent HPLC modules saves benchspace
- Fast analysis in less than 5 minutes.
Interested? Contact Da Vinci Laboratory Solutions for more information.
Digital Edition
PIN 25.5 Oct/Nov 2024
November 2024
Analytical Instrumentation - Picturing Viscosity – How Can a Viscometer or a Rheometer Benefit You? - Sustainable Grease Formulations: Evaluating Key Performance Parameters and Testing Method...
View all digital editions
Events
Dec 03 2024 Dusseldorf, Germany
Dec 08 2024 Anaheim, CA, USA
Turkey & Black Sea Oil and Gas
Dec 11 2024 Istanbul, Turkey
Dec 19 2024 Aurangabad, India
Jan 20 2025 San Diego, CA, USA