• Analysis of Elemental Sulfur in Liquefied Gases

Analytical Instrumentation

Analysis of Elemental Sulfur in Liquefied Gases

Sep 20 2022

To prevent corrosion in refining processes and to comply with environmental regulations it is vital to determine elemental sulfur. Da Vinci Laboratory Solutions (DVLS) has developed a dedicated gas chromatographic (GC) application for the analysis of elemental sulfur in LPG based on the ASTM D7756 configuration.

GC application for Elemental Sulfur & Oily Residue

Elemental sulfur is an unstable component with a tendency to stick to active surfaces, which makes traditional injection techniques challenging. To address this challenge Da Vinci introduced an alternative and fast gas chromatographic (GC) method, which includes a direct injection of the LPG. This direct injection technique ensures full transfer of the elemental sulfur component onto the column.

The GC application uses the Agilent 8890 Gas Chromatograph (GC) configured with the DVLS Pressure Station and Liquefied Gas Injector to safely handle and inject LPG samples. A Flame Ionization Detector and a Sulfur Chemiluminescence Detector are used to simultaneously detect oily residue and elemental Sulfur. This configuration is based on the ASTM D7756 configuration to determine oily residue in LPG within 20 minutes.

The Liquefied Gas Injector (LGI) injects the LPG as a liquid directly into the GC inlet, which eliminates the need for evaporating large volumes of liquefied gases. In a product video the system configuration is explained followed by a demonstration of the sampling & residue analysis in LPG using the Liquefied Gas Injector.

Elemental Sulfur Study

The calibration curve of the elemental sulfur analysis shows a linear correlation for the analysis range of 0.15 – 22 mg/kg. The calibration of five replicates displays a relative standard deviation below 5% at each concentration level.

Next a study of six LPG samples containing elemental sulfur was performed. This study demonstrates that the Sulfur application analyzes the different amounts of elemental sulfur, ranging from 0.1 to 20 mg/kg S. A Limit of Detection (LOD) of 0.01 mg/kg S for individual components was achieved.

Various Applications

Da Vinci introduced the Liquefied Gas Injector in 2010 for the safe, fast and accurate analysis of oily residue in liquefied petroleum gases by gas chromatography. Adapting the hardware setup allows to extend the analysis to a wider application range comprising of:

  • Oily residues and light contaminants in LPG (ASTM D7756, EN 16423)
  • Elemental sulfur in LPG
  • Amines in LPG
  • Nitrogen, carbon dioxide, hydrogen sulfide and hydrocarbons in (un)stabilized gas condensate
  • Composition & impurities analysis in butadiene, crude C4 & raffinate (ASTM D2593, D4424, D2426)
  • Oxygenate traces in liquid hydrocarbon matrices (ASTM D7423, D7754)
  • Benzene en Toluene in LPG (ASTM D7756)

A new brochure describes the various applications of the DVLS Liquefied Gas Injector. Contact us to request for the new brochure of the Liquefied Gas Injector.


Digital Edition

PIN 25.2 Apr/May

May 2024

Safety - Carbon monoxide toxic and flammable gas detection Analytical Instrumentation - Density: A fundamental parameter at critical stages within the petroleum sector - Advancements and...

View all digital editions

Events

The World Biogas Expo 2024

Jul 10 2024 Birmingham, UK

Thailand Oil & Gas Roadshow 2024

Jul 11 2024 Rayong, Thailand

HPLC 2024

Jul 20 2024 Denver, CO, USA

ICMGP 2024

Jul 21 2024 Cape Town, South Africa

Colombia Oil & Gas

Jul 24 2024 Bogata, Colombia

View all events