• Analysis of Manganese in Gasoline by EDXRF
    Rigaku NEX QC⁺ Energy Dispersive X-ray Fluorescence Analyser

Analytical Instrumentation

Analysis of Manganese in Gasoline by EDXRF

The measurement of manganese (Mn) in gasoline is demonstrated.

Background

Methylcyclopentadienyl manganese tricarbonyl (called MMT or MCMT) is an anti-knock agent added to motor gasoline and Avgas to boost octane rating, replacing tetraethyl lead (TEL) in many regions of the world. In motor gasoline the Mn content is typically between 50-500 mg/kg, and can be as high as 3000 mg/kg (approximately 3 g/L) in Avgas. Reliably characterising the Mn content of gasoline ensures optimum engine performance based on the engine’s compression ratio and other geometrical and mechanical operating conditions. To meet the needs of the industry, Rigaku offers the NEX QC+ analyser, a simple and versatile benchtop EDXRF spectrometer for the analysis of manganese in gasoline.

Instrumentation

Model: Rigaku NEX QC+ 
X-ray tube: 4W Ag-anode 
Detector: SDD 
Sample Type: Gasoline 
Film: Chemplex Etnom (3um) 
Analysis Time: 100 sec 
Environment: Air

Sample Preparation

Ensure each sample is homogeneous and stable. Shake sample gently and allow any bubbles to settle. Fill a 32mm XRF sample cup with 4g of sample to ensure consistent sample depth. Etnom® film (3um) is used, which does not degrade quickly with gasoline or oxygenated gasoline. The sample cup lid must be vented with a small hole to ensure vapor pressure does not build. Due to high evaporation rate of gasoline, it is recommended to make the measurement immediately after filling the cup. [Note: Use of internal standard is not required.]

Calibration

Empirical calibration is made using commercially available certified gasoline calibration standards containing the Mn additive. A typical calibration is demonstrated here for the range 25-500 mg/kg Mn.

See TABLE (Calibration)

Precision

Instrument repeatability (precision) is determined by ten repeat analyses of each sample in static position using a 100 sec analysis time per measurement.

See TABLE (Precision)

Detection Limit

The Lower Limit of Detection (LLD) is defined as three times the standard deviation of ten repeat analyses of a blank gasoline sample containing no Mn. The following typical LLDs are reported here using 100 and 300 sec measurement times.

See TABLE (Detection Limit)

Conclusion

The results shown here indicate the Rigaku NEX QC+ EDXRF analyser can be used to reliably measure Mn in gasoline and Avgas without the need to use an internal standard.


Digital Edition

PIN 25.5 Oct/Nov 2024

November 2024

Analytical Instrumentation - Picturing Viscosity – How Can a Viscometer or a Rheometer Benefit You? - Sustainable Grease Formulations: Evaluating Key Performance Parameters and Testing Method...

View all digital editions

Events

Valve World Expo

Dec 03 2024 Dusseldorf, Germany

ASTM - DO2 Meeting

Dec 08 2024 Anaheim, CA, USA

Turkey & Black Sea Oil and Gas

Dec 11 2024 Istanbul, Turkey

IGC 2024

Dec 19 2024 Aurangabad, India

Clean Fuels Conference

Jan 20 2025 San Diego, CA, USA

View all events