Interference-free Sulphur Analysis in Fuels

Aug 18 2014

Author: Dr. Stefan Jezierski on behalf of Analytik Jena GmbH

Free to read

This article has been unlocked and is ready to read.

Download

There is no question that fuels need to be monitored. During the combustion of carbon fuels, not only the carbon dioxide focused on in public discussions, but also other combustion products arise. In addition to various nitrogen oxides, sulphur oxides are of particular interest. If sulphurous organic substances are combusted, SO2 is generated, provided there is sufficient oxygen supply. This compound acts as potent toxin, especially in the catalytic converters found in private cars in industrial nations. Furthermore, sulphuric acid (H2SO3) results from dissolving in water, one of the main causes of acid rain. Especially in Europe, this has caused multiple problems in nature and the environment, agriculture, as well as building damage during the 80s of the 20th century. Thus the intensive control of SO2 entering the environment is of vital interest. 

A major introduction of SO2 is through the combustion of fossil fuels. Whereas large plants, such as power stations, often have flue gas desulphurisation equipment, the emission from smaller contributors, such as heating systems in houses or cars, must be controlled through the fuel itself. The limit values for sulphur content of fuels for transport were significantly reduced in recent years. In the year 2009 a 100 percent market penetration of sulphur-free fuel (< 10 ppm) was implemented in the European Union, and other industrial and industrialising nations are adapting to these limits. [1] Reductions for other fuel types are being planned. The sulphur content for maritime fuels in the North Sea and Baltic Sea, for example, as well as in the English Channel, is currently being reduced to 0.1% (presently 1.5). The same limit values will apply from that date onwards for the US and Canadian coasts. For the remaining EU waterways a limit of 0.5% (presently 4.5%) is being introduced. [2]

The safe monitoring of the sulphur content in various fuels is therefore a basic requirement for laboratories working in the various sectors of petrochemistry or refineries. For this purpose they also need a powerful analysis technology that can analyse both the end products and their raw materials.

For the detection of SO2 in combustion gases, detection using fluorescence induced by ultraviolet light has been proven to be successful. These UVF detectors have the advantage of measuring the SO2 content in a flowing gas across a large measuring range. The advantage of UV fluorescence compared to other detection methods is its high sensitivity, longterm stability and high applicative flexibility. This is countered by the problem of cross-sensitivity to nitrogen monoxide (NO) which has a fluorescence within similar wavelengths.

Experimental

All measurements were performed using an elementary analyser from the multi EA® 5000 series. Dependent on the sample matrix and based on the double furnace option, a horizontal or vertical method could be used. The device also has an integrated gas box for controlling the gas supply 
as required. 

For the horizontal method the multi EA® 5000 (fig. 1) was equipped with an automatic boat feeder. This controls, in combination with a flame sensor, the feed parameters autonomously and independently of the matrix and thus guarantees the complete conversion of the analytes. The combustion takes place in a two-zone combustion tube. During the first phase the analytes are pyrolised in an argon atmosphere. These pyrolysis gases are passed to the second zone of the combustion tube where they are incinerated at 1050°C in a pure oxygen atmosphere. This produces SO2 from sulphurous organic compounds in accordance with the following equation:

O2+R-S SO2+CO2+ H2O

The analysis gases are passed to the UVFD module after drying and purifying. 

The UV fluorescence detector used achieves a measuring range of max. 10,000 ppm and min. 5 ppb in sulphur without having to take recourse to rather lengthy procedures, such as “trap and release”. 

Quantification took place via a calibration using suitable standards. To this end di-benzo-thiophene was dissolved in ultrapure isooctane. From a 1000 ppm stock solution dilutions in isooctane of up to 50 ppb concentrations were prepared. To prepare a calibration graph, these standards were analysed using the same methods as for the subsequent analytes. 

The different fuel types were supplied for combustion either vertically (highly volatile analytes) or horizontally dependent on their viscosity. In both methods the MMS5000 was used for automatic sample supply. Equipped with a syringe, it can dose these liquids. In combination with a gripper the samples can be handled as if they were solids. The control of the analyser and analysis of results took place using the device software.

Sulphur Detection in Fuels

An important requirement for analysis equipment in petrochemistry is the largest possible dynamic measuring range to allow for the analysis of both, raw or intermediate products, often heavy loads and end products with very low concentrations. Table 1 shows an example of this kind of analysis. Two production steps of bio diesel from wood oil were analysed. The raw product had an intensive colour and rather viscous consistency. It was analysed analogue to a solid in the horizontal mode. It was found that the concentration of sulphur in this raw material is rather high with 2400 ppm. The refined end product, on the other hand, a clear liquid of low viscosity, could be measured in a vertical furnace position. The sulphur content of less than 5.5 ppm is clearly below the statutory limit values, so that this product may be used as fuel additive or pure.

The high measuring range of the multi EA® 5000 permits the monitoring of the entire production process and can be used for different matrices arising.

The following table shows the results for different fuel samples. The high precision of the results also means that a reliable result is available after only a few repeat measurements. All Diesel samples were measured horizontally as liquid in this case, the other samples vertically.  

Use of UVF Detectors with a Micro Plasma Optimiser (MPO)

When using UVF detectors, the already mentioned cross-sensitivity to NO can lead to distorted results. The fluorescence of NO tends to only have an intensity of 0.6 – 2% compared to the fluorescence intensity of SO2, but in very nitrogen-rich samples with very low sulphur content this can cause problems. For example: 200 ppm nitrogen lead to a total sulphur result of 1 – 4 ppm. With the very low limit values for fuels the limit value may thus quickly be exceeded as a false positive result.

However, especially in modern fuels, nitrogen compounds are added to improve the combustion properties. Substances, such as amyl nitrate, cyclohexyl nitrite or tri-ethylene glycol nitrite make up up to 5 %-wt, depending on the fuel. These so-called cetane improvers modify the ignitability and octane rating of petrol and diesel positively, but due to the increased nitrogen content ways must be found during the sulphur analysis using UVF detectors to eliminate this irrelevant content. This can be achieved by processing the analytes, e.g. by dilution with pure fuel. Or these expensive procedures are dispensed with and an interference-free detection method is used.

To measure samples with high nitrogen content, a UVF detector with the patented MPO technology was used. The MPO creates plasma in the gas flow that selectively oxidises NO to NO2, so that the combustion products of the nitrogen compounds no longer display a specific fluorescence. This allows for the real sulphur content to be detected even if there is a large mass percentage of nitrogen in the sample, without the need for additional processing steps.

Table 3 shows the sulphur content of different fuels dependent on the activity of an MPO. The detected mass percentage of nitrogen for the respective sample is also stated. This was measured using a chemiluminescence detector (CDL) located after the UVF detector in an additional detector module within the measuring gas flow. The software permits the easy change-over from single element mode with the MPO enabled for sulphur to multiple element mode with the MPO disabled. It becomes apparent that at higher nitrogen concentrations the use of the MPO is useful to derive the true sulphur content. 

Summary

The examples above demonstrate that the multi EA® 5000 is ideally suited as a versatile sulphur analyser for the quick and simple monitoring of intermediate and end products. With its high measuring range and the flexibility of sample supply it permits the analysis of various matrices from highly viscous or solid to highly volatile material mixes. The use of an MPO further offers the option to perform reliable and specific analyses with the UVF detector used even where nitrogen content is high. The change-over from multiple element mode to single element mode is easily possible at any time, allowing the end user to adjust individually to the analytical problem faced.

The use of these methods is not only limited to fuels and their manufacture. There are many applications where a non-negligible amount of nitrogen is present in addition to sulphur in the analysis of organic matrices, such as polymers, grease, lubricants and oils for different uses as well as gases.

Sources 

[1] http://www.ifqc.org/Spotlight.aspx?Id=135 accessed 2014-05-13

[2] a) http://europa.eu/legislation_summaries/other/l21050_de.htm accessed 2014-05-13

[2] b) http://www.imo.org/OurWork/Environment/PollutionPrevention/AirPollution/Pages/Sulphur-oxides-(SOx)-%E2%80%93-Regulation-14.aspx accessed 2014-05-13

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

PIN 25.5 Oct/Nov 2024

November 2024

Analytical Instrumentation - Picturing Viscosity – How Can a Viscometer or a Rheometer Benefit You? - Sustainable Grease Formulations: Evaluating Key Performance Parameters and Testing Method...

View all digital editions

Events

FILTECH

Nov 12 2024 Cologne, Germany

Intech

Nov 12 2024 Tel Aviv, Israel

LPG Week 2024

Nov 18 2024 Rome, Italy

OSEA 2024

Nov 19 2024 Singapore

CIPPE

Nov 19 2024 Shanghai, China

View all events